No.8213-00
Infrared Thermometer with Circlular Laser Marker
SK-8130
JAN code : 4974425821303
Calibration : Available
JCSS Accredited Calibration : Not available
This product is discontinued. Refer to other infrared thermometers.
Feature
■ Equipped with a circular laser marker
■ IPX4 waterproof: resists water splashes from any direction.
■ Backlight on display
■ Emissivity is adjustable between 0.10 and 1.00 in steps of 0.01.
What is emissivity?
- All objects emit infrared rays corresponding to surface temperature. The amount of energy given by radiated infrared rays depends not only on temperature but also material and surface condition of the object, thus the coefficient named emissivity is necessary to calculate temperature.
■ Setting upper and lower limit alarm, "peak" mode (continuous display of maximum), "valley" mode (continuous display of minimum) are available.
When "peak" mode is chosen, SK-8130 measures temperature continuously and shows a maximum value constantly. "Valley" mode does the same for a minimum value.
■ PSC mark is labeled on the device to prove its quality based on Japan's Consumer Product Safety Act.
Description
■ Caution! This device is for industrial use, not for clinical use.
■ Measuring spot
A laser projected from the laser window in the center of the sensing area navigates the measuring area and its center.
■ Measured area depends on distance from the object as shown below.
■ For the better measurement
A square and a circle represent the object and the measured area respectively.
|
|
|
Best | Get close if possible. |
Temperature of other than the object is measured. Get close. |
Specifications
Cat. No. | 8213-00 | |
Product name | Infrared Thermometer with Circlular Laser Marker | |
Model name | SK-8130 | |
Measuring range | −50 to 500°C | |
Accuracy | ±1%rdg (200°C or more) ±2°C (1 to 199°C) ±5°C (0°C or less) under blackbody calibration with ambient temperature 23±5°C, emissivity 0.95 | |
Temperature drift | ±0.1%rdg/°C (200°C and up) ±0.2°C/°C (below 200°C) where emissivity is 0.95 | |
Repeatability | ±1°C (object temperature > 30°C, ambient temperature 23±5°C) | |
Resolution | 1°C | |
Response speed | 90% in 1.5 seconds | |
Emissivity | Adjustable between 0.10 and 1.00 in steps of 0.01 | |
Distance coefficient | D:S = 15:1 (D is distance, S is diameter of spot circle) | |
Detector | Thermopile | |
Measured wavelength | 8 to 14 μm | |
Laser marker | Circle marker Source wavelength: 650 nm Output: Less than 1 mW Certified for PSC, a standard based on Japan's Consumer Product Safety Act, as a class II portable laser device. | |
Functions | Automatic power-off, automatic backlight, Hi/Lo alarm, three modes (ordinary measurement mode, peak mode, valley mode) switchable | |
Alarm volume | 55 dB | |
Waterproofness | IPX4 waterproof | |
Operation ambient | 0 to 50°C, less than 85%rh without condensing | |
Power | AA alkaline battery x2, consumption 120 mA | |
Battery life | Approx. 16 hours if laser is on at room temperature | |
Materials | Main body | Polycarbonate |
Lens | Silicone | |
Dimensions | 44 (W) x 152 (H) x 27.5 (D) mm | |
Weight | 140 g including batteries | |
Standard accessories | Neck strap x1, AA alkaline battery x2, instruction manual |